Troubleshooting
32
Model 515B/C Service Manual Rev. 01
Light interference. Ambient light source (sunlight, warming lights, etc.)
are interfering with sensor operation. Shield sensor from these light
sources. This message will disappear once the problem is corrected.
Sensor fault. Remove sensor from use and contact qualified service per-
sonnel.
Monitor fault. Record the error number that appears in the pulse rate dis-
play (XX will vary, depending upon the fault). Remove the monitor from
use and contact qualified service personnel (report the number that
appears in the pulse rate display).
Bad Signal. Monitor not receiving valid signals from sensor. May be
caused by excessive motion, cardiac arrhythmia or other situations lead-
ing to poor signal. Check patient status, reposition sensor. This message
will disappear once the problem is corrected.